Model:FT-CW32H
Brand:tianyi
Fluorescent Quantitative PCR Detection System Instrument Basic Structure
Fluorescent Quantitative PCR Detection System is a fully integrated fluorescent PCR amplification and data analysis system that uses fluorescent dyes to measure and analyze DNA in real time.Fluorescent Quantitative PCR Detection System is an instrument for real-time detection of reactions. It is mainly composed of a gene amplification thermal circulation system, a fluorescence real-time detection system, a microcircuit control system, a computer and application software. Two of the core functional modules are: the thermal circulation system and the fluorescence real-time detection system. Among them, the working principle of the gene amplification thermal cycle system is basically the same as that of the traditional gene amplification instrument, and the thermal cycle process is completed by semiconductor heating and refrigeration. The fluorescence detection system mainly consists of fluorescence excitation components, optical signal transmission components, fluorescence detection components, and control systems.
Fluorescent Quantitative PCR Detection System Instrument Technical Indicators
The "///" symbol indicates that the national standard currently does not define this indicator or mandatory requirement.
Metric Name | CW16H | CW32H | CW48H | |
Appearance | ||||
External dimensions | 235mm*385mm*175mm (width*deep*height) | |||
weight | 5.6kg | 5.7kg | 5.8Kg | |
Electrical parameters | ~220V/50Hz, 255W | |||
Data interface | USB 2.0 *2 (two on the right) | |||
Environment parameters | ||||
Running conditions | Temperature: 10-30℃, humidity: 20%~80% | |||
Transportation and storage conditions | Temperature: -20~55℃, humidity: 20%~80% | |||
Altitude | <2500 meters | |||
Noise level | A weight, <60dB | |||
Sample parameters | ||||
Sample capacity | 16*0.2mL | 32*0.2mL | 48*0.2mL | |
Test tube type | Single tube, eight-line test tube | |||
Sample volume | 15-100uL | |||
Temperature characteristics | ||||
Heating/cooling method | Semiconductor heating/refrigeration | |||
Temperature range | 4℃-99℃ | |||
Maximum heating rate | ≥3.5℃/s | |||
Average heating rate | ≥2.5℃/s | |||
Maximum cooling rate | ≥3.5℃/s | |||
Average cooling rate | ≥2.5℃/s | |||
Temperature control accuracy | ≤±0.01℃ | |||
Temperature accuracy | ≤±0.1℃ | |||
Temperature uniformity | ≤±0.3℃ | |||
Optical properties | ||||
Number of fluorescence detection channels | 4 channels | |||
Light emitting devices | High brightness LED | |||
Lighting devices | High sensitivity, high signal-to-noise ratio photodiode | |||
Adapt probe or dye | First channel: 470/520 FAM, SYBR Green | |||
Second channel: 530/570 HEX, JOE, VIC | ||||
Third channel: 580/610 ROX, CY3.5, Texas-Red | ||||
Channel 4: 630/670 CY5 | ||||
Detection sensitivity | 1 copy | |||
Linear detection range | 100~1010 copies | |||
Linear correlation coefficient | ≥0.999 | |||
Channel cross-talk | No cross-talk | |||
Detect repeatability | ≤1.0% |
What kinds of Wind Speed Sensors are there on the market and how should consumers choose?There are a wide variety of Wind Speed Sensors available in the market. However, according to the detection principles, they can be classified into mechanical sensors, ultrasonic sensors, thermal sensors, Pitot...
The core advantage of the Wireless Weather Station lies in its integrated composite meteorological sensor. This key component can accurately measure various meteorological parameters, effectively meeting application needs in different scenarios and providing reliable support for obtaining meteorolog...
A Negative Oxygen Ion Monitoring Station is a professional set of equipment used to monitor the number of negative oxygen ions in the air in real-time. It records the activity status of negative oxygen ions in the air through specific sensors, then converts these statuses into readable values via a...
In outdoor natural settings such as dense forests, mountain streams, vast rivers, lakes and seas, and around rushing waterfalls, people often feel the air is exceptionally fresh. The key reason behind this is that the air in these places contains a sufficient amount of negative oxygen ions. Essentia...