Model:FT-SZZD
Brand:tianyi
1.Total Nitrogen Analyzer principle (resorcinol method)
Total Nitrogen Analyzer provides an easy-to-use method for measuring total nitrogen in a variety of liquids, providing a fast and accurate method for total nitrogen analysis.In an alkaline medium at 120°C, potassium persulfate is used as an oxidant.Not only can the ammonia nitrogen and nitrite nitrogen in the water be converted into nitrates, but most of the organic nitrogen is converted into nitrates, and then the absorbance is measured at a specific wavelength with an ultraviolet spectrophotometer.
2.Total Nitrogen Analyzer product features
The new optoelectronic quantitative technology is not affected by interference such as large chromaticity, many suspended objects, and many bubbles.It has high reliability and accuracy and strong anti-interference ability.
High detection accuracy, low measurement limit, and low long-term drift of the instrument.
It has one-click self-test function and self-diagnosis function.
It adopts a single-channel high-integrated valve set, making maintenance and cleaning simple and fast.
All-weather networking function, monitoring the operating status of the instrument anytime and anywhere.
It has alarm function, quality control function and counter control function.
It has automatic cleaning function and automatic calibration function.
It has full-point measurement, interval measurement and external control measurement.
It works automatically after power failure recovery.
Automatic prompts and automatic reset functions for faults and reagents (samples).
It has fault logging function.
Can save historical data for more than 3 years.
Satisfies the separation of cleaning liquid and waste liquid
3.Total Nitrogen Analyzer Technical Parameters
Measurement method: UV spectrophotometry for alkaline potassium persulfate digestion
Measuring range: 0-5 mg/L; 0-20 mg/L; 0-50 mg/L; range can be customized
Display error: ≤±5% detection limit: 0.01mg/L
Repeatability: ≤±3% Waste liquid volume: 6mL/time
Memory effect: ≤±1%Real water sample comparison: ±10%
Constant temperature time: 900 seconds Voltage stability: ≤±5%
Measurement cycle: 51 minutes Calibration cycle: any specified time
Maintenance cycle: ≥720h/time Maintenance workload: <2 hours/month
Reagent consumption: 3 months/500ml display output: configured with a 10.1-inch color LCD touch screen
Color rendering temperature: The recommended temperature is 125℃, which can be set according to the actual water sample condition
Data export: The measured value can be imported into the USB drive and saved in the USB drive.
Signal output: RS485/RS232/USB interface/standard 2-channel 4-20mA output/standard two-channel switching input and output
Environmental requirements: Indoors with controllable temperature, recommended temperature (5-28℃), humidity ≦90% (no condensation)
Power supply and power: (220V±22) V/AC, (50±0.5Hz), 5A, 150W
Instrument size: upper cabinet 600*450*300mm; upper cabinet 700*450*300mm
Although traditional mechanical wind speed and direction meters have, to some extent, met basic requirements, they have limitations in terms of accuracy, response speed, and reliability. With the advancement of technology, the 3D Ultrasonic Anemometer based on the time - difference method has been d...
Agriculture, as the foundation of human survival, is intricately linked to the weather. Weather conditions profoundly influence every aspect of agricultural production. During the sowing stage, seed germination requires suitable accumulated temperature, typically above 10°C. If the temp...
Negative oxygen ion monitoring plays a crucial role in ecological environment research and air quality management. Negative oxygen ions are negatively charged ions in the air, which are widely distributed in the natural environment. Their concentration levels are closely related to air quality. High...
People have been measuring the wind for a long time, and the tools for wind measurement have been constantly upgraded. In the 15th century, Leonardo da Vinci designed a vane anemometer, marking the beginning of the era of scientific wind measurement. In 1846, Thomas Romney Robinson, a British naval...